You are seeing this message because your Web browser does not support basic Web standards. Find out more about why this message is appearing and what you can do to make your experience on this site better.


ABOUT ARCHIVES
Advanced Search

Welcome   | My Account | E-mail Alerts | Access Rights | Sign In


  Vol. 60 No. 10, October 2003 TABLE OF CONTENTS
  Archives
  •  Online Features
  Basic Science Seminars in Neurology
 This Article
 •Full text
 •PDF
 • Reply to article
 •Send to a friend
 • Save in My Folder
 •Save to citation manager
 •Permissions
 Citing Articles
 •Citing articles on Web of Science (12)
 •Contact me when this article is cited
 Related Content
 •Similar articles in this journal
 Topic Collections
 •Neurology, Other
 •Alert me on articles by topic
 Social Bookmarking
  Add to CiteULike Add to Connotea Add to Del.icio.us Add to Digg Add to Reddit Add to Technorati Add to Twitter What's this?

The Application of Implant Technology for Cybernetic Systems

Kevin Warwick, DSc; Mark Gasson, BSc; Benjamin Hutt, PhD; Iain Goodhew, BEng; Peter Kyberd, PhD; Brian Andrews, PhD; Peter Teddy, FRCS; Amjad Shad, FRCS(SN)

Arch Neurol. 2003;60:1369-1373.

Objective  To assess the usefulness, compatibility, and long-term operability of a microelectrode array into the median nerve of the left arm of a healthy volunteer, including perception of feedback stimulation and operation of an instrumented prosthetic hand.

Setting  The study was carried out from March 14 through June 18, 2002, in England and the United States.

Results  The blindfolded subject received feedback information, obtained from force and slip sensors on the prosthetic hand, and subsequently used the implanted device to control the hand by applying an appropriate force to grip an unseen object. Operability was also demonstrated remotely via the Internet, with the subject in New York, NY, and the prosthetic hand in Reading, England. Finally, the subject was able to control an electric wheelchair, via decoded signals from the implant device, to select the direction of travel by opening and closing his hand. The implantation did not result in infection or any perceivable loss of hand sensation or motion control. The implant was finally extracted because of mechanical fatigue of the percutaneous connection. Further testing after extraction has not indicated any measurable long-term defects in the subject.

Conclusions  This implant may allow recipients to have abilities they would otherwise not possess. The response to stimulation improved considerably during the trial, suggesting that the subject learned to process the incoming information more effectively.


From the Department of Cybernetics, University of Reading, Reading, England (Drs Warwick, Hutt, Kyberd, and Andrews and Messrs Gasson and Goodhew); and the Department of Neurosurgery and Neurosciences, The Radcliffe Infirmary, Oxford, England (Drs Teddy and Shad).